Title Analysis of Lotus japonicus
Description Leaves of Lotus japonicus lines (MG-20 and B-129) were analyzed by LC-FT-ICR-MS. Two Arabidopsis thaliana samples were used as references.
Authors Takeshi Ara, Hideyuki Suzuki, Daisuke Shibata, Nozomu Sakurai, Ryosuke Sasaki, Makoto Hirosawa, Lab. Genome Biotech, Kazusa DNA Research Institute, Japan; Center for Development of Biotech Industries, Kazusa DNA Research Institute, Japan
Reference Direct submission
Comment NS deposited the metadata, most of which were retrieved from MassBase. For more detailed information, please contact the authors directly or via NS (email: metabolonote [at] kazusa.or.jp).

Sample Information

ID S1
Title Lotus japonicus B-129 Leaf
Organism - Scientific Name Lotus japonicus
Organism - ID NCBI Taxonomy:34305
Compound - ID
Compound - Source
Preparation
Sample Preparation Details ID
Comment

Analytical Method Information

ID M2
Title LC-FTICR-MS, ESI Positive analysis
Method Details ID MS1
Sample Amount
Comment [MassBase ID] MDLC1_05759


Link icon database.png Link icon massbase.png

The raw (binary) and near-raw (text) files of this analysis are available at MassBase.

Analytical Method Details Information

ID MS1
Title LC-FTICR-MS, ESI Positive analysis
Instrument LC-FT-ICR-MS; [Agilent 1100 LC (Agilent Technologies); Finnigan LTQ-FT (Thermo Fisher Scientific K.K.)]
Instrument Type LC-FTICR-MS
Ionization ESI
Ion Mode Positive
Description Harvested sample is frozen by liquid N2 and resulting powder (100mg) are solved in 300uL 80% methanol solution. 20uL sample is injected into HPLC after 0.2um membrane filter treatment. HPLC conditions: Agilent 1100 series (Agilent), Solvent: A; 0.1% formic acid aq. B; ACN (addition 0.1% formic acid fc.), Gradient: (B);3 to 40% (0.0 to 23.0 min), 40 to 95% (23.0 to 23.1 min), 95% (23.1 to 25.0 min), 3% (25.1 to 30.0 min), Column temp.: 30 degree C, Flow rate=0.5mL/min, PDA: 200-650 nm (2 nm step). FT-ICR-MS conditions: Filter 1: FTMS + c norm !corona !pi res=50000 o(200.0-1500.0), Rejected mass=266.0000;294.0000;391.0000.
Comment_of_details [MassBase Method ID] MTLC0349

Data Analysis Information

ID D1
Title Auto peak detection by PowerGetBatch
Data Analysis Details ID DS1
Recommended decimal places of m/z
Comment


Data Analysis Details Information

ID DS1
Title Auto peak detection by PowerGetBatch
Description
Comment_of_details
Personal tools
View and Edit Metadata
Variants
Views
Actions