SE196:/S22/M01

From Metabolonote
jump-to-nav Jump to: navigation, search

Sample Set Information

ID SE196
Title A lipidome atlas in MS-DIAL 4
Description We formulated mass spectral fragmentations of lipids across 117 lipid subclasses and included ion mobility tandem mass spectrometry (MS/MS) to provide a comprehensive lipidome atlas with retention time, collision cross section and MS/MS information. Using human, murine, algal, and plant biological samples, we annotated and semi-quantified 8,051 lipids by MS-DIAL 4 (http://prime.psc.riken.jp/) with a 1–2% estimated false discovery rate, using the lipidomics standards initiative level 2 or 3 definitions.
Authors Hiroshi Tsugawa, Kazutaka Ikeda, Mikiko Takahashi, Aya Satoh, Yoshifumi Mori, Haruki Uchino, Nobuyuki Okahashi, Yutaka Yamada, Ipputa Tada, Paolo Bonini, Yasuhiro Higashi, Yozo Okazaki, Zhiwei Zhou, Zheng-Jiang Zhu, Jeremy Koelmel, Tomas Cajka, Oliver Fiehn, Kazuki Saito, Masanori Arita, and Makoto Arita
Reference Tsugawa, H. et al. A lipidome atlas in MS-DIAL 4. Nat Biotechnol (2020). https://doi.org/10.1038/s41587-020-0531-2
Comment


Link icon database.png Link icon dropmet.png

The raw data files are available at DROP Met web site in PRIMe database of RIKEN.

Link icon article.png

Sample Information

ID S22
Title 22_Mouse_Lung (RIKEN IMS Sciex Mouse tissues)
Organism - Scientific Name Lung (Mouse)
Organism - ID NCBI taxonomy 10090
Compound - ID
Compound - Source
Preparation
Sample Preparation Details ID
Comment

Analytical Method Information

ID M01
Title RIKEN IMS Sciex Mouse tissues: Lung (Mouse) LC-ESI-MS analysis
Method Details ID MS02
Sample Amount Positive mode: 1 microLiter; Negative mode: 2 microLiter
Comment


Analytical Method Details Information

ID MS02
Title Method 2: SCIEX TripleTOF DDA normal mass range method
Instrument LC, Waters Acquity UPLC system; MS, AB Sciex TripleTOF 5600+ or 6600 system
Instrument Type UPLC-QTOF-MS
Ionization ESI
Ion Mode Positive and Negative
Description Methanol, isopropanol, and acetonitrile of LC-MS grade were purchased from Wako. Ammonium acetate and EDTA were purchased from Wako and Dojindo, respectively. Milli-Q water was purchased from Millipore. SPLASH Lipidomics I or EquiSPLASH used as internal standards were purchased from Avanti Polar Lipids.

<LC Method>

The LC system consisted of a Waters Acquity UPLC system. Lipids were separated on an Acquity UPLC Peptide BEH C18 column (50 x 2.1 mm; 1.7 µm) (Waters, Milford, MA, USA). The column was maintained at 45°C at a flow-rate of 0.3 mL/min. The mobile phases consisted of (A) 1:1:3 (v/v/v) acetonitrile:methanol:water with ammonium formate (5 mM) and 10 nM EDTA and (B) 100% isopropanol with ammonium formate (5 mM) and 10 nM EDTA. A sample volume of 0.5-3 µL, which depended on biological samples, was used for the injection. The separation was conducted under the following gradient: 0 min 0% (B); 1 min 0% (B); 5 min 40% (B); 7.5 min 64% (B); 12 min 64% (B); 12.5 min 82.5% (B); 19 min 85% (B); 20 min 95% (B); 20.1 min 0% (B); and 25 min 0% (B). Sample temperature was maintained at 4°C.

<MS Method>

Mass spectrometric detection of lipids was performed on a quadrupole/time-of-flight mass spectrometer TripleTOF 5600 or 6600 (SCIEX, Framingham, MA, USA). All analyses were performed at the high resolution mode in MS1 (~35,000 full width at half maximum (FWHM)) and at the high sensitivity mode (~20,000 FWHM) in MS2. Data dependent MS/MS acquisition (DDA) was used. The parameters were MS1 and MS2 mass ranges, m/z 70-1250; MS1 accumulation time, 250 ms; MS2 accumulation time, 100 ms; collision energy, +40/-42 eV; collision energy spread, 15 eV; cycle time, 1300 ms; curtain gas, 30; ion source gas 1, 40(+)/50(-); ion source gas 2, 80(+)/50(-); temperature, 250°C(+)/300°C(-); ion spray voltage floating, +5.5/-4.5 kV; declustering potential, 80 V. The other DDA parameters were dependent product ion scan number, 16; intensity threshold, 100 cps; exclusion time of precursor ion, 0s; mass tolerance, 20 ppm; ignore peaks, within m/z 200; and dynamic background subtraction, True. The mass calibration was automatically performed using an APCI positive/negative calibration solution via a calibration delivery system (CDS).

Comment_of_details