Browse wiki

jump-to-nav Jump to: navigation, search
SE131:/MS1
MS Comment of details Sawada et al. Plant Cell Physiol. 2009 Jan;50(1):37-47
MS Description The UPLC (Waters) conditions were manually The UPLC (Waters) conditions were manually optimized based on the separation patterns of 12 methionine-derived glucosinolates and were as follows: fl ow rate 0.24 ml min –1 ; solvents A, 0.1% formic acid in water and B, 0.1% formic acid in acetonitrile; gradient program of B (0 min, 0%; 0.25 min, 0%; 0.4 min, 9%; 0.8 min, 17%; 1.9 min, 100%; 2.1 min, 100%; 2.11 min, 0%); 3 min cycles with a temperature of 38°C.The TQMS detection conditions were the same as those for FIA-TQMS, except that the source temperature was 130°C. ept that the source temperature was 130°C.
MS ID MS1  +
MS Instrument HPLC: Waters Acquity UPLC system; MS: Waters Q-TOF Premier  +
MS Instrument Type UPLC-QTOF-MS  +
MS Ion Mode Positive and Negative  +
MS Ionization ESI  +
MS Title LC-ESI-QqQ-MS  +
Modification dateThis property is a special property in this wiki. 4 April 2018 08:23:37  +